In line with Philips commitment to providing customer choice, IntelliVue is compatible with a range of SpO2 technologies and sensors from Philips, Masimo, and Nellcor.
Toggle Navigation. Sign in. My Bench Sign out. One moment please My Bench Close. Sign In. Not A Member? Join MedWrench. Receive Summary Emails? Receive Instant Notices. Display in Community Feed. You can adjust your Community Subscriptions in Settings.
Remove Subscription Confirm Remove. Follow Us. Request a quote. Request Parts. Add to My Bench. AG Biomedical. Envi Health Solutions. Integris Equipment. Med Repair Tech. Your monitor comes with a set of preconfigured Screens, optimized for common monitoring scenarios such as OR adult, or ICU neonatal.
A Screen defines the overall selection, size and position of waves, numerics and SmartKeys on the monitor screen when you switch on. You can easily switch between different Screens during monitoring.
Screens do NOT affect alarm settings, patient category and so forth. When you switch from a complex to a less complex Screen layout, some measurements may not be visible but are still monitored in the background. If you do not want to change the entire Screen content, but only some parts of it, you can substitute individual waves, numerics, high-res waves, or trends. Be aware that these changes cannot be stored permanently in Monitoring Mode.
Change HiResTrend , and then select the wave or numeric you want, or select the highresolution trend wave you want from the list of available waves. If you do not see Change Numeric in the menu, this Screen may be configured to always display the numeric beside its wave. Changing the wave will automatically change the numeric.
In the Change Screen menu, the changed Screen is shown linked to the original Screen and marked with an asterisk. The ten most recently-used Screens including up to three modified Screens can be accessed.
Modified Screens are still available in the. If a modified Screen was the last active Screen when the monitor was switched off, it is retained unless Automat. Default is configured to Yes.
A second display, showing the same Screen as the main display, can be connected to any of the monitors, for viewing only. The MP90 can have two main displays and additionally, as a third main display, the D80 Intelligent Display connected. All main displays can be configured and operated individually using standard input devices. For MP90 monitors with multiple displays and multiple input devices, the usage and behavior can be configured according to specific requirements at installation for example, use for two independent operators or tracking of mouse input across two displays.
For details refer to the Service Guide. When two operators are using two displays, the scope of an action depends on the type of operation:. If you are operating two displays with one remote SpeedPoint, to navigate from one display to another:. The content of each Screen can be changed individually as described in the previous section. If you are operating two or three displays, you can choose Screens for all displays from one location:.
Select Profiles in the monitor info line of the first display,. Select Display 1 , Display 2, or Display 3 then select the Screen you want to display. When two displays are mounted next to each other or one above the other, a special Screen can be assigned which spans across both displays. The Screen content for these Tall and Wide Screens can then use the increased area available with two displays.
Certain windows for example: cardiac output procedure can only be shown on one display at a time. If you try to open one of these windows when it is already shown on another display, you will see a blank grey window with a cross through it.
If a visitor Screen is configured for your monitor, you can use it to clear the screen of all waves and numerics but continue to monitor the patient with active alarms and trend storage at the bedside and Information Center.
You can change the name of the visitor Screen in Configuration Mode. Profiles are predefined monitor configurations. They let you change the configuration of the whole monitor so you can adapt it to different monitoring situations.
The changes that occur when you change a complete profile are more far reaching than those made when you change a Screen. Screens affect only what is shown on the display.
Profiles affect all monitor and measurement settings. The settings that are defined by Profiles are grouped into three categories. These categories are:. If you are using a second display, each display can have its own individual screen selection. When you change the profile, the screen selection configured for the new profile becomes active. These relate directly. Changes you make to any element within the settings blocks are not saved when you discharge the patient, unless you save them in Configuration Mode.
You might find it helpful to think of the three categories in terms of a restaurant menu. The Measurement Settings are like the dessert course. From these you build your meal. Depending on your monitor configuration, when you switch on or discharge a patient the monitor either continues with the previous profile, or resets to the default profile configured for that monitor.
WARNING If you switch to a different profile, the patient category and paced status normally change to the setting specified in the new profile. However some profiles may be setup to leave the patient category and paced status unchanged.
Always check the patient category, paced status, and all alarms and settings, when you change profiles. Settings or. Monitor Settings to call up a list of the settings blocks in each category. Your monitor has a default profile that it uses when you leave Demonstration, or Service modes, or when you discharge a patient. This profile is indicated by a diamond. Some profiles are locked, so that you cannot change them, even in Configuration Mode.
These are indicated by this lock symbol. Each aspect of how the monitor works and looks is defined by a setting. There are a number of different categories of settings, including,. Screen Settings , to define the selection and appearance of elements on each individual Screen. Measurement settings , to define settings unique to each measurement, for example, high and low alarm limits. Monitor settings , including settings that affect more than one measurement or Screen and define general aspects of how the monitor works, for example, alarm volume, reports and recordings, and display brightness.
All settings are reset to the stored defaults:. Default is set to. Each measurement has a setup menu in which you can adjust all of its settings. You can enter a setup menu:. Then select the measurement name from the popup list. With this SmartKey you can access any setup menu in the monitor. This guide always describes the entry method using the setup menu. But you can use any method you prefer.
The monitor stops data acquisition and alarming for this measurement. A measurement automatically switches off if you disconnect its module or measurement server. If you disconnect a transducer, the monitor replaces the measurement numeric with question marks. The screen display indicates the active setting. For some measurements, such as EEG, you can choose which numerics to view on the screen.
This displays the measurement Wave menu, which has only waverelated measurement settings. Lowering the wave speed compresses the wave and lets you view a longer time period. Increasing the speed expands the waveform, giving you a more detailed view.
To set an individual channel speed, select a numeric value from the list of available speeds. This overrides the wave group speed setting and set the speed for the individual wave channel on the monitor Screen. The wave channel speed is independent of the wave label depicted in the channel, if you change the wave, the new wave will retain the set channel speed. You can freeze waves on the screen and measure parts of the wave using cursors. The waves are frozen with a history of 20 seconds so that you can go back and measure what you have seen.
If you are using touch, this automatically positions the vertical cursor. The cursor can be repositioned by touching the required point on the wave, or. The vertical cursor moves through the time axis and the current value is displayed next to the cursor. The horizontal cursor measures the wave value, which is displayed above the cursor line.
You can measure multiple invasive pressures, temperatures, and SpO 2 simultaneously. The monitor uses labels to distinguish between them. The default settings defined in the profile such as measurement color, wave scale, and alarm settings are stored within each label. When you assign a label to a measurement, the monitor automatically applies these default settings to the measurement. The labels assigned are used throughout the monitor, in reports, recordings, and in trends.
Your monitor may be configured to use a Restricted or Full label set. The Full label set provides extra labels for Pressure and Temp.
See the sections in this Instructions for Use on Pressure and Temp for more information. This may cause a label conflict with other monitored measurements. Be aware that connecting a monitor using the Full label set to an Information Center with certain software revisions may affect the availability of measurement information from the additional labels on the Information Center. To change a measurement label of a measurement with multiple labels invasive pressure, temperature, or SpO 2 ,.
The monitor automatically applies the scale, color, etc. Now you want to use the same module to measure ABP with a new patient. When you connect the pressure transducer to the module, the pressure shown on the screen still uses the CVP color and wave scale and is labeled CVP.
To rectify this, just change the pressure label to ABP. Now the pressure has the correct color, the wave is shown in the correct scale, and the appropriate alarm limits for ABP are active. Each label must be unique, that is, it can only be assigned once. If you need to use two identical pressures, you must assign different labels to them, for example, P and ICP. Measurement labels are stored in the measurement device module or measurement server.
If you try to use two measurement devices that have identical labels, this causes a label conflict in the monitor. All the currently available measurement devices are depicted in the Measurement Selection window. Any measurement labels causing a label conflict are shown in red. Welcome to ManualMachine. We have sent a verification link to to complete your registration.
Log In Sign Up. Forgot password? Enter your email address and check your inbox. Please check your email for further instructions. Enter a new password. Procedure Window Accessing the Setup C. Measurements Documenting C. Measurements C. Introducing the IntelliVue Family The Philips IntelliVue family of patient monitors offers a monitoring solution optimized for the surgical, cardiac, medical and neonatal care environments.
Remote Alarm Device The Remote Alarm Device provides audio and visual indicators of alarms, in addition to those shown on the display.
Two color coded alarm lamps right-hand lamp flashes red or yellow for patient alarms, left-hand lamp flashes light blue for INOPs Alarms off lamp - when illuminated it indicates that all alarms are deactivated. Power on LED - green when monitor is on. Flexible Module Server MA.
You might have a version of the MMS that 3 does not have this connector. Cardiac Output orange; Temperature connectors brown optional. Operating and Navigating Everything you need to operate the monitor is contained on its screen.
Pause duration depends on monitor change screen menu configuration. Select again to immediately re-enable alarm indicators. Operating and Navigating 1 Basic Operation Selecting Screen Elements Select a screen element to tell the monitor to carry out the actions linked to the element. Using the Setup Menu. Behavior follows the Silence permanent key configuration. Behavior follows the Pause Alarms permanent key configuration.
Exact behavior depends on Pause Alarms permanent key configuration 3 Main Screen - closes all open menus and windows and return to the main screen. Using a Mouse or Trackball If you are using a mouse or trackball, select screen elements by clicking on them press and release the left mouse button. Using Keys The monitor has four different types of keys. Permanent Keys A permanent key is a graphical key that remains on the screen all the time to give you fast access to functions.
Main Setup - enter main setup menu. SmartKeys A SmartKey is a configurable graphical key, located at the bottom of the main screen.
Operating and Navigating 1 Basic Operation start a delayed recording Realtime Record SmartKey to access pop- up recording keys set wide automatic alarm limits set narrow automatic alarm limits access wedge procedure window access the Loops window review vital signs trend review graph trend access event surveillance access calculations access the calculator access the Drug Calculator gas analyzer - exit standby mode suppress zero for all gas measurements unpair equipment access ST Map application start Lead Capture only available access remote applications if Application if Information Center is connected Server is connected access EEG CSA access the EEG montage display VueLink information access timers access the spirometry data window switch CO 2 pump off Hardkeys A hardkey is a physical key on a monitoring device, such as the zero pressure key on the MMS or a setup key on a module.
Pop-Up Keys Pop-up keys are task-related graphical keys that appear automatically on the monitor screen when required. Using the On-Screen Keyboard Use this as you would a conventional keyboard.
Using the On-Screen Calculator You can use the on-screen calculator to perform any of the standard operations for which you would normally use a handheld calculator. To change to a different mode: 1 Select the Main Setup menu. Config When the monitor is in Demonstration Mode, Configuration Mode, or Service Mode, this is indicated by a box with the mode name in the center of the Screen and in the bottom right-hand corner. Standby Mode Standby mode can be used when you want to temporarily interrupt monitoring.
Understanding Screens Your monitor comes with a set of preconfigured Screens, optimized for common monitoring scenarios such as OR adult, or ICU neonatal. To change the selection of elements on a Screen, 1 Select the element you want to change. The changed Screen is shown with an asterisk in the monitor info line.
Using Additional Displays In the Change Screen menu, the changed Screen is shown linked to the original Screen and marked with an asterisk. Up to three modified Screens can be accessed via the Change Screen menu.
Modified Screens are still available in the Change Screen menu. Using Additional Displays A second display, showing the same Screen as the main display, can be connected to any of the monitors, for viewing only.
If you are operating two displays with one remote SpeedPoint, to navigate from one display to another: 1 Move the highlight to the Main Screen key and then turn one click further. Using the Visitor Screen If a visitor Screen is configured for your monitor, you can use it to clear the screen of all waves and numerics but continue to monitor the patient with active alarms and trend storage at the bedside and Information Center.
Understanding Profiles Profiles are predefined monitor configurations. When you leave Demonstration Mode, the monitor uses the default profile. Settings or Monitor Settings to call up a list of the settings blocks in each category. Default Profile Your monitor has a default profile that it uses when you leave Demonstration, or Service modes, or when you discharge a patient. Locked Profiles Some profiles are locked, so that you cannot change them, even in Configuration Mode.
Understanding Settings Each aspect of how the monitor works and looks is defined by a setting. There are a number of different categories of settings, including, Screen Settings , to define the selection and appearance of elements on each individual Screen Measurement settings , to define settings unique to each measurement, for example, high and low alarm limits Monitor settings , including settings that affect more than one measurement or Screen and define general aspects of how the monitor works, for example, alarm volume, reports and recordings, and display brightness.
Default is set to Yes. Changing Measurement Settings Each measurement has a setup menu in which you can adjust all of its settings. Changing Wave Speeds Lowering the wave speed compresses the wave and lets you view a longer time period. Changing the Wave Group Speed The wave speed group setting defines the speed of all the waves in the group.
Changing Wave Speed for a Channel To change the wave speed of an individual wave channel, 1 Enter the Wave menu for a measurement by selecting its wave. Freezing Waves You can freeze waves on the screen and measure parts of the wave using cursors. Freezing An Individual Wave To freeze a wave, 1 Enter the Wave menu for the measurement by selecting the wave on the screen. The realtime wave is replaced with the frozen wave.
All realtime waves are replaced with frozen waves. Measuring Frozen Waves To measure a frozen wave, 1 Select the frozen wave. To change the wave speed: 1 Select the frozen wave. Updating The Frozen Wave To update the wave, that is freeze the current wave: 1 Select the frozen wave. Releasing Frozen Waves To release frozen waves, 1 Select a frozen wave. All frozen waves are released. Using Labels You can measure multiple invasive pressures, temperatures, and SpO 2 simultaneously. Changing Measurement Labels e.
Pressure To change a measurement label of a measurement with multiple labels invasive pressure, temperature, or SpO 2 ,.